– Large LCD display waveform and data
– High-resolution thermal printing array system
– Frequency response is as high as 150 Hz
– Capable of printing continuously on channel trace and annotations including lead mark and parameter such as sensitivity, paper speed and filter operation status
– High resolution digital filter inhibits baseline drift without affecting ECG waveform
– Floating input circuit meets IEC safety standard
– Rechargeable battery supporting more than 50 patient examinations
– Internal software: GB, FR, IT, ES, DE, TR
– Manual: GB, IT (ES on request)
Technical Specifications
ECG instrument Input
• Safety Type: Class I, type CF applied part, which has defibrillation-proof applied part
• Lead Standard 12 leads
• Input way: Floating and defibrillation protection
• Patient leak current: <10µA • Sampling rate: 1000Hz • A/D conversion: 12bit • Frequency response 0.05Hz~150Hz (-3dB, 0.4dB) • Time constant: >3.2s
• CMRR: >60dB, >100dB(Add filter)
• Enduring polarization: voltage ±500mV
• Noise level: <15µVp-p
Dispose Wave Data
• EMG interference filter 35Hz (-3dB)
• AC filter 50Hz/60Hz (=20dB)
Waveform output
• Recording way: Thermal printing system
• Recording paper: 50mm(W)×20m(L)
• Sensitivity: 5, 10, 20mm/mV
• Paper speed:
– Auto-record: auto leads changing
– Rhythm record: 25 mm/s, 50 mm/s
– Manual-record: 25 mm/mV, 50 mm/mV
Power supply
• AC: 110-220V, 50/60 Hz
• DC: 7.4V, 3700 mAh rechargeable lithium polymer battery
Fuse
• 2 pcs f5×20mm AC time lag; T250mA/250V(Power Supply:220V)
Form
• Size 315 x 215 x 77 mm
• Net Weight 2.25 kg
Standard accessories
• ECG cable – 1 set
• Limb electrode clamp – 1 set
• chest electrodes – 1 set
• Ground wire
• Power connection
• Paper roll