HF electrosurgical unit suitable to perform middle monopolar or bipolar electrosurgery. By selection of the modes of operation it provides pure cut, coagulated cut, forced coagulation, soft coagulation and, with forceps, bipolar coagulation. The use of either split or normal neutral plate is allowed. MB 160 is recommended in case of damp operational fields, for instance mucous, with consequent need to have reserve of usable power. For these reasons, MB 160 is suggested for Gynecological and orthopedic surgery.
Leading-Edge Technology This advanced concept device enables highly professional surgery intervention thanks to ergonomic and safety features derived from high power theatre apparatus.
Open Circuit (OC) Alarm warns the user that connection into the neutral plate has not been made or that the cable is disconnected or the neutral plate is not in contact with the patient. The alarm is signalled by an intermittent beep together with a flashing red light on the front panel. In this case HF emission for cut and coagulation operation is switched off.
OVT Alarm. The unit includes an internal circuit protection warning when the current delivery lasts too much.
Technical Specifications
• Minimum preselectable power: 0
• Level step: 1
• Maximum output power CUT (W): 120 – 250 ohm
• Maximum output power BLEND (W): 90 – 200 ohm
• Maximum output power COAG FORCED (W): 80 – 150 ohm
• Maximum output power COAC SOFT (W): 60 – 100 ohm
• Maximum output power BIPOLAR (W): 40 – 100 ohm
• Crest Factor CUT: 1.5
• Crest Factor BLEND: 2.1
• Crest Factor COAG FORCED: 2.0
• Crest Factor COAG SOFT: 1.7
• Crest Factor BIPOLAR: 1.5
• Working frequency: 600 kHz
• Weight: 7 kg
• Size LxHxD mm: 260 x 110 x 265
• Operating voltage (switchable): 115-230
• Mains frequency (Hz): 50-60
• Fuses 230Vac (5×20) TIMED: 2x 3.15A
• Fuses 115Vac (5×20) TIMED: 2x 6.3A
• Electrical input power (VA): 300
• Electrical input current (230Vac) (A): 1.3
• Electrical input current (115Vac) (A): 2.6
• Five steps adjustable sound level (from 55- to 75dBA)
• Self-check
• Power accuracy output warning
• Split or not split patient plate allowed
• Last working condition storing
• Electrical Class (EN60601 – 1): I CF
• MDD 93/42/EEC Class: II B
• Patient circuit: F
• Conform to EN60601-1-4 (1998)
Standard accessories
• MB single-use two button handles
• MB operative foot-switch 30577 (usable as alternative to handle)
• MB flat neutral plate Power cable 2 m
• Kit of 10 short autoclavable electrodes, Ø 2.4 mm
• User manual: available in English, Spanish, Italian, French, Polish